Factor 01Wavelengths
More wavelengths are not automatically better.
The light output of a panel is limited. The more wavelengths are used, the more the available output is spread across those wavelengths.
So the right question is not “how many?” but “by what criteria were they selected?” Three criteria are decisive:
Criterion 01
CCO activation
The absorption curve of cytochrome c oxidase has peaks and troughs. 830 nm sits at a peak. Placing light on a weak part of the curve costs energy without delivering a comparable return.
Criterion 02
Body of studies
There is a reason why 660 nm appears in more than 400 studies, while neighbouring wavelengths have been studied in only a fraction of that number. Weak evidence makes the data sheet longer, but does not automatically make the result better.
Criterion 03
Penetration depth
Red light reaches about 1-3 mm, near-infrared about 3-7 mm. A device that covers only one of these two depths can accordingly reach only one type of application.
Why it matters
Red light works through absorption, not heat. Between 600 and 900 nm – the therapeutic window – light reaches the enzyme cytochrome c oxidase in the mitochondria, where it increases ATP production.
Outside this range, less of use happens: short-wave light is scattered more strongly in the top few millimetres. Infrared above 1000 nm is absorbed by water and mainly produces heat.
How to recognise a good device
- The manufacturer gives specific figures and not just “red and infrared”.
- They can explain why exactly these wavelengths were chosen, instead of just showing the longest possible list.
- They state the manufacturing tolerance in nanometres. Anyone who cannot give one has not measured their LEDs accordingly.
- Both should make you sceptical: only two wavelengths, or a conspicuously large number of wavelengths with no justification.
Market standard vs. heilys®
Most panels use 660 and 850 nm. The reason is simple: these LED types were originally mass-produced for plant lighting and are therefore inexpensive. This is the obvious choice, but not automatically the best-founded one.
heilys® uses six wavelengths with a manufacturing tolerance of ± 5 nm. Together they cover around 82 % of the CCO absorption curve (Tran et al., 2024).
Six is not a number chosen just because it sounds good. It is the result of the three filters mentioned.
| Wavelength |
Depth |
Studies |
Why it is included |
| 630 nm |
1–3 mm |
200+ |
Skin, collagen formation, anti-inflammatory effects |
| 660 nm |
1–3 mm |
400+ |
the most extensively studied wavelength of all |
| 670 nm |
1–3 mm |
150+ |
ATP production, retina, neuroprotection |
| 810 nm |
3–7 mm |
300+ |
deep tissue, muscles, brain |
| 830 nm |
3–7 mm |
250+ |
Peak of CCO activation |
| 850 nm |
3–7 mm |
180+ |
Tendons, joints, immune modulation |
→ Read more: wavelength and its therapeutic effect
Factor 02Control
Precise enough for research. Simple enough for every day.
Two requirements meet here that can easily get in each other’s way. Nevertheless, both have to be met.
The first: precision
Photobiomodulation is a young field of research. A device with a fixed factory setting reflects the state of research in the year it was produced – and stays at that level permanently.
Then there is the individual factor: skin type, build, age, goal and tolerance all differ. Only if settings can be readjusted and saved can you find out what works for your own use.
The second: simplicity
This is where many technically good devices fail. If you first have to go through three menu levels to start a session, it will quickly be skipped after a long working day.
Red light is used regularly over a period of weeks. The user interface does not determine the effect of the light itself, but it very much determines whether the device is used regularly at all.
Market standard vs. heilys®
Standard panels are often hard-wired: one switch, everything on. At the other end of the spectrum are devices with such overloaded apps that in the end only a single mode gets used anyway.
For precision: each of the six wavelengths is individually dimmable. In addition, there are 6 well-founded presets, 12 freely configurable programmes and lifelong software updates.
For everyday use: touch display, app and a remote control with a quick-start button. One press is all it takes to start the most recently saved setting.
A good device solves both separately
Full control when you need it – and one button when you don’t.
The test question: How many steps do I need to start my usual session? If it is more than one, ask yourself: will I really do this every day?
Factor 03Irradiance
Power in reserve makes sense – but only if you can also reduce the intensity. A high maximum output is of little use if it cannot be adapted to the application.
The fallacy
Light acts in the cell via a signal. When light hits cytochrome c oxidase, a small and controlled amount of reactive oxygen species (ROS) is briefly produced. It is precisely this signal that triggers the adaptation: more ATP, better cellular respiration and regeneration.
If the intensity is too high, the cell produces more ROS than its antioxidant systems can neutralise. The signal turns into oxidative stress. Instead of ramping up, the cell then switches into a protective mode. Excess energy can do harm where regeneration was actually meant to be supported. Excessive intensity can therefore reverse the desired effect.
| Application |
Sensible irradiance |
| Skin, anti-ageing, supporting wound healing |
20–50 mW/cm² |
| Muscles, joints, deep tissue |
70–100 mW/cm² |
| Target range for an all-round device |
40–100 mW/cm² |
| over 100 mW/cm² |
no proven additional benefit, but excess ROS and heat |
For context: the midday sun delivers around 100 mW/cm² in total. Of this, about 20-30 mW/cm² lies within the therapeutic window. A device advertised at 200 mW/cm² therefore delivers many times what nature provides in this range.
The measuring trick that almost nobody explains
heilys® Gen300 · measured at a distance of 15 cm
82 mW/cm²
with spectrometer
182 mW/cm²
with Solarmeter
Both values have been measured correctly. However, only one of them is meaningful for the therapeutic effect!
Why reserve still matters
This does not mean, however, that a weak device is sufficient. Irradiance decreases significantly as the distance increases. A panel that still reaches 30 mW/cm² at 15 cm delivers only a fraction of that output at 50 cm. That is why output and treatment distance must always be considered together.
Power in reserve makes sense – but only if you can also reduce the intensity. A high maximum output is of little use if it cannot be adapted to the application.
How to recognise a good device
-
Measured at what distance? Without a distance, the figure is meaningless.
-
Measured with what? Spectrometer or Solarmeter – the difference can be around a factor of 2.
-
Measured or calculated? Real values are often 25-50 % below calculated figures.
-
And most importantly: Does the supplier simply advertise the highest possible number, or do they explain why their value is where it is?
Market standard vs. heilys®
We measure with a spectrometer, state the distance and also publish the Solarmeter value. This lets you compare our figure with other suppliers, even though it makes our number look smaller.
The 82 mW/cm² is not a compromise on the low side but a deliberate choice: enough reserve for different distances without unnecessarily overtaxing the cells.
→ Read more: the right irradiance
Factor 04Dosage
Why too much achieves as little as too little.
Irradiance describes how fast energy arrives. The dose, on the other hand, describes how much energy arrives in total.
Dose (J/cm²) = irradiance (mW/cm²) × time (s) ÷ 1,000
What happens with too little
The cells receive light, but the amount remains below the threshold at which enzyme activity changes measurably. The effect is then not simply weaker – it may not occur at all.
This is a common reason why people give up after a few weeks: too great a distance, sessions that are too short or a device that is too weak.
What happens with too much
The same mechanism is at work here as with excessive intensity, only over time: the ROS signal can tip over into oxidative stress. The cell then scales down instead of up.
Simply doubling the time therefore does not automatically produce a greater effect – in the worst case, it may even produce less.
What this means in practice
The effective range lies between a dose that is too low and one that is too high. Whether the dose is outside this range usually cannot be felt or seen directly during the session.
That is why time and distance are not just non-binding recommendations but part of the application.
Getting started: in the first week, 2-3 min per area every other day. Then increase to 3-5 sessions per week, with a maximum of 20 min per area.
| Goal |
Dose |
Distance |
Duration |
| Skin, anti-ageing |
3–15 J/cm² |
45–60 cm |
approx. 8 min |
| Joint discomfort |
10–40 J/cm² |
15–30 cm |
approx. 10 min |
| Muscle recovery |
20–50 J/cm² |
15–30 cm |
approx. 12 min |
| Energy and vitality |
10–30 J/cm² |
30–60 cm |
approx. 10 min |
| Wound healing support |
10–40 J/cm² |
15 cm |
approx. 7 min |
| Sleep and recovery |
3–15 J/cm² |
30–45 cm |
approx. 7 min |
| Eyes (670 nm) |
3–15 J/cm² |
30 cm |
1–5 min, in the morning |
Irradiance decreases as distance increases. The times and distances given should therefore be considered together. The 82 mW/cm² of the Gen300 refers to a measuring distance of 15 cm.
→ Read more: dosage in red light therapy
Factor 05Homogeneity
A dose figure is only as good as the evenness of the illumination.
A panel consists of many individual LEDs. If the light is not distributed evenly, one area of skin receives considerably more energy than the one next to it. Then not every area receives the same dose – and the calculation from factor 04 no longer applies to the entire surface.
Approach A · panels via the lenses
- There needs to be enough distance between the device and the skin for the optics to determine the distribution.
- Red is meant to act over a wide area, near-infrared to be more tightly focused into depth. If the same optics are used for both, one of the two functions is inevitably not designed appropriately.
-
Test question: Do red and infrared LEDs have different beam angles?
Approach B · wearables via LED spacing
- A mask sits directly on the skin. Over this short distance, no lens can fully even out the individual cones of light.
- Here, what matters most is how closely the LEDs are spaced. The larger the gap, the more pronounced the pattern of bright and dark spots.
-
Test question: How large is the distance between two LEDs?
The test without a measuring device
-
The wall test: point the panel at a light-coloured surface at the usual treatment distance. Visible dots or a honeycomb pattern indicate hotspots.
-
The hand test: hold your hand in front of it at 15-20 cm. Localised warmth suggests uneven illumination.
heilys® Gen panels
Two optics: 60° for the red wavelengths (630/660/670 nm) and 30° for the near-infrared wavelengths (810/830/850 nm).
In the Gen300, 306 LEDs are arranged so that the light cones of both optics overlap. Red and infrared light reach the same area of skin at the same time – without gaps or hotspots.
The dual-lens design is not an extra, but the prerequisite for six wavelengths to work together effectively.
heilys® FaceMask 2.0
A different lever is used here. Because the mask sits directly on the face, the distance between the individual LEDs has been reduced.
More closely spaced points of light produce significantly more even illumination at short distances.
The goal remains the same: in the end, every area of the skin should receive the same dose.
→ Read more: what is the beam angle?
Factor 06Light quality
How clean the light is depends on the driver.
Three properties all depend on the same component – the power supply – and yet hardly appear on any data sheet.
WHY “FLICKER-FREE” DOES NOT ALWAYS MEAN THE SAME THING
Flicker is a visual effect: the brightness of the light fluctuates, usually in time with the mains frequency. Above a certain frequency, you no longer perceive it consciously. The visual system still registers the fluctuation, however. Sensitive people may react to it with restlessness, eye strain or headaches.
The key marketing point here
The term “flicker-free” is used very loosely. Many manufacturers simply mean that the frequency is so high that the flicker lies above the threshold of conscious perception.
But invisible flicker is still flicker. A higher frequency does not eliminate the fluctuation in brightness; it merely hides it. Genuine freedom from flicker is only possible with a DC driver. This supplies the LEDs constantly instead of switching them on and off in cycles. The result is 0 Hz – not just a low value, but no flicker frequency at all.
The decisive question is therefore not just “Is it flicker-free?”, because almost everyone answers yes to that. It is: “Does the device run on direct current, and what is the flicker frequency in hertz?”
PULSED VISIBLE LIGHT AND PHOTOSENSITIVITY
Pulsing is deliberately generated flicker. The LEDs are switched at a fixed frequency, usually between 10 and 50 Hz.
Arguments in favour
- At the same average output, the heat input is lower because the LED can cool down between pulses.
- For individual frequencies around 10 Hz, there are indications of specific effects in neural tissue.
Arguments against
- There is as yet no robust evidence that pulsed light is fundamentally superior to continuous light.
- Calculating the dose becomes more difficult because the duty cycle (= the duty cycle indicates how long the LED is actually switched on within a pulse) also has to be taken into account.
And one point that is more serious than both of the above
There is visible light that pulses at 10 to 50 Hz. This is precisely the frequency range that can trigger seizures in photosensitive people. With a large panel that takes up a large part of the field of vision, this is not a purely theoretical risk. Pulsed visible red light therefore has no place in a panel.
The situation is different with small devices that sit directly on a specific area of the skin: there, the light source is on the skin and not in the field of vision.
Electromagnetic fields
Whether electromagnetic fields are produced is not the decisive question – every power supply produces them. What matters is how much the device adds to the electromagnetic fields already present in the environment.
Why it matters
Every power supply generates an electromagnetic field. During red light sessions you often sit 15 to 60 cm away from the device – and frequently every day.
We expressly do not derive any health claim from this. For us, a different benchmark makes sense.
The right benchmark
Every living space already has an electromagnetic background level from the household wiring, appliances and power supplies.
A well-designed red light device should not significantly raise this existing level. This is technically feasible. And what is feasible should also be implemented.
Market standard vs. heilys®
heilys® Gen panels work with DC drivers. The flicker frequency is 0 Hz.
Pulsing is only possible in the NIR spectrum and can be selected in the app between 10 and 50 Hz. Near-infrared is invisible, so there is no risk of photosensitive reactions from visible pulsed light. The visible red wavelengths in the panel always run continuously – deliberately and without exception.
As for EMF, our values at the point of use are at the level of the normal background level of a home environment. The device therefore adds practically nothing to its surroundings.
→ Read more: pulsed vs. continuous light therapy
Factor 07RELIABILITY & VERIFICATION
What counts is the output in minute ten.
Why it matters
LEDs lose light output when they get too hot. They also age faster. Your dose calculation may then be correct in the first minute, but no longer later on.
Passive: silent and less expensive, but limited in the amount of heat it can dissipate. This makes sense for small devices.
Active: technically more complex and not completely silent, but keeps the output stable. Above a certain power class, active cooling is practically the only option – a 480-watt panel cannot be cooled passively in any sensible way.
How to recognise a good device
-
Does the cooling match the stated output? A high-performance panel without a fan is either not as powerful as claimed or gets hot.
-
How loud is the device? Ideally with a figure in decibels at the usual treatment distance.
- What are the LED lifespan and guarantee?
-
Has anyone tested it independently? The manufacturer’s own measurements are normal; external tests are much rarer.
Market standard vs. heilys®
We use active cooling where the output requires it: in all Gen panels and FullBody panels as well as in the Gen30 and Gen50. The smaller devices manage with passive cooling.
The fans run quietly: approx. 36 dB at 15 cm and approx. 33 dB at 50 cm – roughly in the range of a whisper.
100,000 hours LED lifespan · 5-year guarantee · 30-day right of return · CE, RoHS, WEEE DE 19301058 · developed and tested in Germany.
1,1
Independently tested
The heilys® Gen300 was rated 1.1 by the German Institute for Product Testing (Deutsches Institut für Produkttests). View the test report →